Simplify calibration with hot-swappable smart probes, minimizing logger downtime and maximizing efficiency.
| Logger Compatibility | CP100K:
|
| Temperature Measurement Range | CP100K and CP110K: -40 °C to +99 °C (-40 °F to +210 °F) CP10S and CP11S: -90 °C to +40 °C (-130 °F to +104 °F) |
| Temperature Operating Range, Probe End | CP100K and CP110K: -40 °C to +99 °C (-40 °F to +210 °F) CP10S and CP11S: -90 °C to +40 °C (-130 °F to +104 °F) |
| Temperature Operating Range, Signal Conditioning End | CP100K and CP110K: -20 °C to +70 °C (-4 °F to +158 °F) CP10S and CP11S: -30 °C to +60 °C (-22 °F to +140 °F) |
| Rated Temperature Reading Accuracy | CP100K and CP110K: Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4 °F to +104 °F), typically ±0.3 °C (0.6 °F) Better than ±0.8 °C (±1.4 °F) for -40 °C to -20 °C (-40 °F to -4 °F), typically ±0.5 °C (0.9 °F) Better than ±0.8 °C (±1.4 °F) for +40 °C to +70 °C (+104 °F to +158 °F), typically ±0.7 °C (1.3 °F) Better than ±1.2 °C (±2.2 °F) for +70 °C to +99 °C (+158 °F to +210 °F), typically ±1 °C (1.8 °F) CP10S and CP11S: Better than ±0.5 °C (±0.9 °F) for -20 °C to +10 °C (-4 °F to +50 °F) Better than ±0.7 °C (±1.3 °F) for -50 °C to -20 °C (-58 °F to -4 °F) Better than ±0.7 °C (±1.3 °F) for +10 °C to +30 °C (+50 °F to +86 °F) Better than ±1 °C (±1.8 °F) for -80 °C to -50 °C (-112 °F to -58 °F) Better than ±1 °C (±1.8 °F) for +30 °C to +40 °C (+86 °F to +104 °F) Better than ±1.8 °C (±3.2 °F) for -90 °C to -80 °C (-130 °F to -112 °F) Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration. |
| Temperature Resolution | CP100K and CP110K: 0.1 °C (0.2 °F) for measurements -40 °C to +50 °C (-40 °F to +122 °F) 0.2 °C (0.4 °F) for measurements +50 °C to +70 °C (+122 °F to +158 °F) 0.3 °C (0.5 °F) for measurements +70 °C to +80 °C (+158 °F to +176 °F) 0.4 °C to 0.6 °C (0.7 °F to 1 °F) for measurements +80 °C to +99 °C (+176 °F to +210 °F) CP10S and CP11S: 0.1 °C (0.2 °F) for measurements -80 °C to 0 °C (-112 °F to +32 °F) 0.2 °C (0.4 °F) for measurements 0 °C to +18 °C (+32 °F to +64.4 °F) 0.2 °C (0.4 °F) for measurements −90 °C to −80 °C (−130 °F to −112 °F) 0.3 °C (0.5 °F) for measurements +18 °C to +30 °C (+64.4 °F to +86 °F) 0.5 °C (0.9 °F) for measurements +30 °C to +40 °C (+86 °F to +104 °F) LogTag® Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag® |
| Acquisition Interval | Determined by logger, shortest acquisition interval 30 seconds |
| Sensor Reaction Time | Based on tip, same as for current probes loggers (Low temperature loggers) |
| Environmental | Probe tip: IP68 Probe tip/cable junction: IP64 Signal conditioning: IP44 |
| Power Source | CR2032 3V LiMnO2 Battery (distributor replaceable) |
| Connection Interface | Single wire digital interface to compatible loggers |
| Battery Life | 1 year storage, followed by 3 years of normal use (based on 5 minute logging) |
| Weight | 34g |
| Signal Conditioning Case Material | Polycarbonate |
| Tip Material | Stainless Steel 316 |
| Probe Tip Length | CP100K and CP110K: 65mm exposed tip CP10S and CP11S: 22.5mm exposed tip |
| Probe Diameter | CP100K and CP110K: 3.2mm CP10S and CP11S: 5mm |
| Cable Material | PTFE |