The UTRED30-16 is a two-channel temperature data logger with a large LCD screen and a memory capacity of 32,000 values (16,000 per channel). The device quickly displays temperature fluctuations for the day and the previous 29 days, and issues alerts with an X and sound when the set threshold is exceeded. It is suitable for cold chain monitoring in the storage of vaccines, pharmaceuticals, and food requiring high accuracy and compliance with CDC, FDA, VFC, NIH, PHAC, AAP standards, etc.
| Product Model | UTRED30-16 |
| External Sensor Measurement Range | -40°C to +99°C (-40°F to +210°F) |
| Operating Temperature Range | -30°C to +60°C 0 – 99 % RH (non-condensing) |
| Storage Temperature Range | -10°C to +40°C (32°F to +104°F) |
| Rated Temperature Resolution | Better than 0.1°C (0.2°F) for -30°C to +60°C (-22°F to +140°F) Better than 0.2°C (0.4°F) for -40°C to -30°C (-40°F to -22°F) Better than 0.5°C (0.9°F) for +60°C to +99°C (+140°F to +210°F) LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag® |
| Rated Temperature Reading Accuracy | Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F) Better than ±0.8°C (±1.4°F) for -40°C to -20°C (-40°F to -4°F), typically ±0.5°C (0.9°F) Better than ±0.8°C (±1.4°F) for +40°C to +70°C (+104°F to +158°F), typically ±0.7°C (1.3°F) Better than ±1.2°C (±2.2°F) for +70°C to +99°C (+158°F to +210°F), typically ±1°C (1.8°F) |
| Probe Connectivity | Up to 2 |
| Probe Compatibility | All ST100 precision thermistor external probes can be interchangeably connected to the UTRED30-16 Please refer to the Accessories section for compatible External Probes. Different probe models can be used for each channel. Not compatible with ST10 sensors |
| Sensor Reaction Time | Typically less than 2 minutes(T90) in moving air (1m/s) for ST100S type |
| Recording Capacity | 32,000 total log memory. Provides 16,000 temperature readings per channel Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics |
| Sampling Interval | Configurable from 30 seconds to 18 hours. (Typically 5 minute logging) |
| Alarms | Up to 6 individually configurable alarms for each channel Possible to set alarms separately for Channel 1 and Channel 2 |
| Logging Start Options | Push button start or specific date & time. Optional start delay |
| Download Time | Typically with full memory (32,000 readings) in less than 20 seconds |
| Environmental | IP61 (when vertically mounted or hung with USB dust-cap in place) |
| Power Source | 5V via micro USB socket or 2 x 1.5V AAA batteries. CR2032 LiMnO2 coin cell (backup logging function only) |
| Battery Life | AAA battery up to 12 months – if unit is operated without power supply |
| Real Time Clock | Built-in real time clock Typical accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day) Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
| Connection Interface | USB 2.0 / Micro USB Socket |
| Standard Remote Probe Cable Lengths | Standard: 1.5m (4’11”) Extended: 3m (9’10”) (recommended maximum) |
| Remote Probe Cable Type | PTFE (FDA food contact rated) coaxial |
| Size | 141mm(H) x 72mm(W) x 18mm(T) |
| Weight | 156g with AAA batteries, (132g – excluding probe and AAA batteries) |
| Case Material | Polycarbonate |
User Manual for UTRED 30-16 | |
English Version | VietNamese Version |