TREX 8 have Interchangeable external probe with high-quality gold-plated connectors.
8000-read memory capacity with up to 3 years of battery life.
'OK' and 'ALERT' indicator lights with customizable high and low alert levels.
Industry-best download time – under 5 seconds for a full memory.
| Product Model | TREX-8 |
| External Sensor Measurement Range | -40 °C to +99 °C (-40 °F to +210 °F) |
| Operating Temperature Range | -40 °C to +85 °C (-40 °F to +185 °F) |
| Storage Temperature Range | -10 °C to +55 °C (14 °F to +131 °F) |
| Rated Temperature Reading Accuracy | Better than ±0.5 °C (±0.9 °F) for -10 °C to +40 °C (+14 °F to +104 °F) Better than ±0.7 °C (±1.3 °F) for -30 °C to -10 °C (-22 °F to +14 °F) Better than ±0.7 °C (±1.3 °F) for +40°C to +60°C (+104 °F to +140 °F) Better than ±0.8 °C (±1.4 °F) for -40 °C to -30 °C (-40 °F to -22 °F) Better than ±0.8 °C (±1.4 °F) for +60°C to +80°C (+140 °F to +176 °F) Better than ±1.0 °C (±1.8 °F) for +80 °C to +99 °C (+176 °F to +210 °F) Actual performance is typically much better than the rated values Accuracy figures can be improved by recalibration |
| Rated Temperature Reading Resolution | Less than 0.1 °C (0.2 °F) for measurements -40 °C to +40 °C (-40 °F to +104 °F) Less than 0.2 °C (0.4 °F) for measurements +40 °C to +80 °C (+104 °F to +176 °F) Less than 0.6 °C (0.5 °F) for measurements +80 °C to +99 °C (+176 °F to +210 °F) LogTag Analyzer® currently displays to one decimal place of °C or °F |
| Sensor Reaction Time | Typically less than 2 minutes (T90) in moving air (1m/s) for ST100T, ST100H, and ST100S types |
| Recording Capacity | 8,031 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging |
| Sampling Interval | Configurable from 30 seconds to 18 hours |
| Logging Start Options | Push button start or specific date and time. Optional start delay of up to 72 hours |
| Recording Indication | Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator |
| Download Time | Typically less than 5 seconds for full memory (8,031 readings), depending on computer or readout device used |
| Environmental | IP61 |
| Power Source | CR2450 3V LiMnO2 Battery (Fixed) |
| Battery Life | 1 year storage, followed by 2 – 3 years of normal use (based on 15 minute logging, download data monthly) |
| Real Time Clock | Built-in real time clock Rated accuracy ±25ppm @ 25 °C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (i.e. typically +/- 0.00294 seconds/day/°C) |
| Standard Remote Sensor Cable Length | Standard: 1.5m (4’11″) Extended: 3m (9’10″) (Recommended maximum) |
| Remote Sensor Cable Type | PTFE (FDA food contact rated), coaxial |
| Size | 86mm(H) x 54.5mm(W) x 8.6mm(T) |
| Weight | 36g (without External Probe) |
| Case Material | Polycarbonate |
TREX-8 User Manual | |
English Version | VietNamese Version |