LogTag® Vaxtag® is a specialized temperature monitoring device for vaccine refrigerators in healthcare facilities and intermediate storage areas, complying with WHO PQS standards. The device features a display and data logging capabilities, allowing for quick viewing of temperature and time statistics over several days. When the temperature exceeds the set threshold, the device displays a visual alert. Data can be viewed directly on the screen or downloaded via the LogTag® Analyzer software for analysis, storage, and management.
| Product Model | TRID30-7FW (Vaxtag) |
| Sensor Measurement Range | -30 °C to +60 °C (-22 °F to +140 °F) |
| Operating Temperature Range | -30 °C to +60 °C (-22 °F to +140 °F) |
| Storage Temperature Range | -10 °C to +40 °C (-14° F to +104 °F) |
| Rated Temperature Reading Accuracy | Better than ±0.5 °C (±0.9 °F) for measurements from -20 °C to +40 °C (-4 °F to +104 °F), typically ±0.3 °C (0.6 °F) Better than ±0.8 °C (±1.5 °F) for other measurements - typically ±0.5 °C (0.9 °F) Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration |
| Rated Temperature Reading Resolution | 0.1 °C (0.2 °F) for measurements -30 °C to +40 °C (-22 °F to +104 °F) 0.2 °C (0.4 °F) for measurements above +40 °C (+104 °F) LogTag® Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag® |
| Sensor Reaction Time | Typically less than 5 minutes (T90) in moving air (1m/s) |
| Recording Capacity | 7,770 temperature readings. 32 days @ 6 minute logging. Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values |
| Sampling Interval | Factory set to 6 minute interval |
| Logging Start Options | Factory set to Push button start |
| Recording Indication | State indicator “RECORDING” |
| Alarm Triggering | Low alarm: Exposure to a single temperature event of -0.5 °C or below for 60 minutes. High alarm: Exposure to a single temperature event of +8 °C or above for 10 hours. |
| Download Time | Typically less than 5 seconds for full memory, depending on computer or readout device used |
| Environmental | IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung |
| Power Source | CR2450 3V LiMnO2 (Fixed Battery) |
| Battery Life | 1 year storage, followed by 2 years of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly) |
| Real Time Clock | Built-in real time clock Rated accuracy ±25ppm @ 25 °C (equivalent to 2.5 seconds/day) Rated temperature coefficient is -0.034 ±0.006ppm/°C (i.e. typically +/- 0.00294 seconds/day/°C). |
| Size | 93mm(H) x 54.5mm(W) x 8.6mm(T) |
| Weight | 43g |
| Case Material | Polycarbonate |
VAXTAG Device User Guide | |
Enghlish Version | VietNamese Version |